Assemble your CORE Digital Attachment

mceclip0.png

Follow our 3 step process to attach the CORE to your analog stethoscope.

  1. Detach the Chest Piece
    • Remove the chest piece from your analog stethoscope. Grip the chest piece with one hand and the tubing in the other and pull them apart. This may require some force. 
  2. Install CORE and Chestpiece 
    • Select either the standard or pediatric size tubing connector and insert your chest piece into the tubing. If the standard tubing connector feels loose on the chestpiece, please try the pediatric tubing connector instead.

    • Carefully screw in the chestpiece and tubing connector into the end of the CORE that’s closest to the power toggle. Please make sure the threads are aligned when you are screwing it in otherwise it may be prone to cross-threading.

    • mceclip1.png
  3.  Attach CORE to Stethoscope Tubing 
    • Select the corresponding barb (either pediatric or regular) and carefully screw it into the other end of the CORE. Please make sure the threads are aligned when you are screwing it in otherwise it may be prone to cross-threading. 

    • On the other end of the CORE, select the corresponding barb (either pediatric or regular) and screw it into the base of the CORE. Please make sure the threads are aligned when you are screwing it in otherwise it may be prone to cross-threading.

    • At this time, you can connect the barb to the stethoscope tubing. 
    • mceclip2.png

Power On/Off

Power On

CORE_Power_ON.gif

Depress the power slider to move the switch from the OFF to the ON position. 

“ON” is when the toggle is flushed with the surface of the volume buttons. You will see a white blinking light to indicate that the CORE has been turned ON. You can now hear amplified sounds through the stethoscope.

Power Off

CORE_power_Off.gif

OFF is when the toggle is protruding from the surface of the volume buttons. The white light will disappear and you can now only hear analog sounds through the stethoscope earpieces. 

Standby Mode

The CORE has a 5 minute standby mode to conserve battery life. The white light will disappear when the CORE goes into standby mode and the toggle will remain protruding from the surface of the volume buttons. To power the device back ON, push the toggle to the OFF and ON position. The white light will appear to indicate the device is ON. 

Adjust the Volume

CORE’s sound level can be amplified in 7 increments up to 40X amplification of an acoustic stethoscope when listening with the stethoscope earpieces. Change the volume level by clicking the plus (+) and minus (-) volume buttons on the side of CORE.

If you are using headphones to listen to the sounds, please adjust your phone’s volume to change the CORE’s volume. 

Note: Changing the volume on the CORE does not affect the volume of a recording on the App. Recording volume must be adjusted in the app or on the phone itself.

Charge your CORE

When the yellow light starts blinking on the CORE, the battery in CORE will need to be charged. 

Insert the included micro USB cable into the USB port on the device and plug the other end into a UL-certified USB wall charger. The LED will turn solid yellow, signifying that it is charging. The LED will change to solid green when the device is fully charged.

To fully charge the battery, it should take about 2.5 hours. When fully charged, the CORE should last for about 8 hours of continuous use. We recommend turning the device OFF between each use to maximize use of the battery.

NOTE: CORE will not turn on while it is plugged in and charging.

Clean your CORE device

The stethoscope and CORE should be cleaned between each patient's use. All cleaning instructions pertaining to the original stethoscope apply.

Under normal conditions it is unnecessary to remove CORE from the stethoscope tubing for cleaning. All external parts of the hardware can be cleaned with 70% isopropyl alcohol wipes.

NOTE: DO NOT immerse the device in any liquid or subject it to any high-pressure/autoclave sterilization processes.

If it becomes necessary to remove CORE, pull the stethoscope tubing off of the metal stem on both ends of the device. Wipe all parts of the stethoscope clean with 70% isopropyl alcohol wipes including CORE’s surface, stethoscope tubing, tubing connector, and chest piece. Reassemble the stethoscope by reinserting the metal stems into the stethoscope tubing as before.

Ensure the best audio quality

Please follow the steps below to achieve the best quality auscultation when using your CORE and stethoscope earpieces:

  1. If possible, perform auscultation in a quiet environment. While the CORE has active noise cancellation, there is a chance that some environmental noise will still be audible during auscultation. 
  2. Use the bell to listen to low frequency sounds like heart sounds. Use the diaphragm to listen to high frequency sounds like lung sounds.
  3. Hold the diaphragm firmly against the surface you are auscultating. Pressing too tightly against the surface can muffle sounds.

LBL 252, Rev A, April 28, 2021.